At Hingmed, we developed the DBP-01P as a fully automatic upper-arm blood pressure monitor for hospitals, clinics, community health centers, pharmacies and other medical establishments. Our goal is to provide a simple, efficient workflow for measuring adult systolic blood pressure, diastolic blood pressure and pulse rate.
The DBP-01P uses the oscillometric measurement method. After each measurement, it displays the results on a clear LED screen. When the printing function is enabled, the built-in thermal printer also produces an immediate paper record for review or filing.
Although we designed the DBP-01P for convenient one-button operation and self-service screening, it remains a clinical measurement device. A blood pressure reading should be interpreted by a healthcare professional and should not be treated as a diagnosis on its own.
What Makes Our DBP-01P Different
Our DBP-01P is equipped with a wide-range cuff suitable for arm circumferences from 17 to 42 cm. The cuff can be used with either the left or right arm and adjusted in different directions to accommodate a variety of users and operating positions.
We have also incorporated an elbow-positioning sensor into the device. This feature helps guide the user’s arm into the correct position before measurement begins. Correct positioning reduces avoidable measurement failures and makes the monitor more suitable for self-service environments.
Voice guidance provides instructions during operation and can announce the completed reading. The built-in printer uses 58 mm thermal paper to create a clear record of the measurement for immediate review.
Our double-pump system is designed to support faster operation and a longer product life. Pulse-wave motion-tolerance technology helps the device obtain readings when slight movement occurs, although users should still remain quiet and motionless during measurement. Other practical features include low-noise operation, an antibacterial cuff cover and an optional disposable sleeve that can support infection-control procedures.
We use a modular design to make maintenance and component replacement more convenient. The compact automatic blood pressure module can also be incorporated into a larger healthcare kiosk or telemedicine system.
For system connectivity, our DBP-01P product page lists USB, RS-232 and SD-card ports, with Ethernet available as an option. We can also provide an SDK and technical support for system integration, including electronic patient record applications. The security and regulatory compliance of an integrated solution will depend on the complete data system and the purchaser’s implementation.
Preparing the DBP-01P for Use
Before operating the monitor, please read our DBP-01 series user manual, including its warnings, precautions and maintenance instructions.
Place the DBP-01P on a stable surface. Do not install it in a location exposed to water, direct sunlight, excessive heat, moisture, dust, vibration or impact. It should also be kept away from stored chemicals, corrosive substances and areas where combustible gas may be generated.
The specified operating environment is a temperature of 5°C to 40°C and relative humidity of 10% to 95%. Connect the supplied power cable to a 110–240 V AC, 50/60 Hz power source. Always use a properly grounded three-prong outlet because the device is classified as Class I equipment for protection against electric shock.
Before switching on the monitor, inspect the casing, screen, cuff, buttons and power cable. Do not use the device if you find visible damage, moisture, significant contamination or an abnormal electrical connection.
For models using the printing function, press the printer-cover button and open the cover. Insert the 58 mm thermal paper into the paper compartment. Pull the leading edge of the paper through the outlet, and then close the cover while leaving a short section exposed.
After switching on the DBP-01P, confirm that the display operates normally. Check the Start/Stop button, emergency-stop function and printer before beginning routine measurements.
Confirming That the User Is Suitable
Our DBP-01P is intended to measure blood pressure and pulse rate in adults in hospitals and other medical establishments.
Do not use the device to measure an arm receiving an intravenous infusion or blood transfusion. Applying cuff pressure to that arm may create a safety risk.
Common arrhythmias can also prevent an automatic blood pressure monitor from achieving its stated performance. Users with anticoagulant conditions or blood-coagulation disorders may experience congestion where the cuff is applied, even when the arm is positioned correctly. Healthcare professionals should apply the appropriate clinical protocol in these situations.
Taking a Blood Pressure Measurement
Before starting, sit comfortably with your back straight. Relax, remain quiet and avoid unnecessary movement. Use a bare arm whenever possible. Thin clothing may be acceptable, but thick clothing around the upper arm can interfere with cuff contact and cause measurement errors.
Our DBP-01P can measure either arm. The current user manual expresses a preference for the right arm. Insert your arm fully into the cuff sleeve and place it on the wrist support.
Position your elbow directly over and against the elbow pad button. This step is important because the positioning sensor must detect the elbow before the device can complete a normal measurement. If the elbow switch is not pressed, the monitor may display an EC19 error.
Once your arm is properly positioned, press the Start/Stop button. The device will automatically close the cuff around your arm, inflate it and apply the required pressure.
Remain still while the measurement is in progress. Do not speak, move your arm or change your sitting position. Body movement, conversation, an incorrectly positioned arm, a rough or unstable surface and external vibration can all affect a blood pressure reading.
When the measurement is finished, the DBP-01P automatically releases the cuff pressure and returns the cuff to its original state. Your systolic pressure, diastolic pressure and pulse rate will appear on the LED display.
Depending on the selected settings, the device can announce the results and print them automatically. The paper record can then be reviewed by a healthcare professional or processed through the medical establishment’s configured information system.
If another measurement is required, wait at least two to three minutes before repeating the procedure.

Stopping a Measurement
If you feel uncomfortable or need to interrupt the process, press the Start/Stop button. The device will stop measuring and quickly release the cuff pressure.
Use the emergency-stop control if an abnormal operating condition occurs. If the cuff does not inflate and release normally within 2.5 minutes, remove your arm manually. Prolonged cuff inflation may obstruct blood flow and cause discomfort.
Never continue using the monitor when it produces abnormal sounds, smoke, a burning smell or repeated system errors. Switch off the power and arrange an inspection by authorized service personnel.
Responding to Error Codes
An error code identifies an operating or measurement problem that should be addressed before another reading is attempted.
EC01 may indicate a loose cuff or an arm circumference outside the supported 17–42 cm range. Check the position of the arm and its contact with the cuff.
EC04 indicates a weak signal, which may result from a weak pulse or poor cuff contact. Reposition the arm and try again.
EC06 means excessive movement or signal interference occurred during measurement. Remain quiet and still before repeating the procedure.
EC19 indicates that the arm position is incorrect or the elbow switch has not been pressed. Reposition the elbow directly over the positioning button.
EC64 means that the printer is out of paper. Install a new roll of 58 mm thermal paper. EC65 indicates that the printer cover is not properly closed.
Persistent air-circuit, valve, motor, communication or system faults should be handled by authorized service personnel. Users should not open or modify the internal components.
Understanding the Results
The DBP-01P has a specified blood pressure accuracy of ±3 mmHg. Its specified pulse-rate accuracy is ±3 beats per minute or ±3%, whichever is greater.
The systolic measurement range is 40–260 mmHg, while the diastolic range is 20–210 mmHg. The supported pulse-rate range is 40–200 beats per minute.
These performance specifications do not mean that one measurement can diagnose a medical condition. Blood pressure varies with posture, physical condition, recent activity and other factors. Unexpected, unusually high, unusually low or inconsistent readings should be assessed under the medical establishment’s clinical procedures.
Cleaning and Maintaining Our DBP-01P
Before cleaning, switch off the monitor and disconnect it from AC power. Wipe patient-contact surfaces with a clean, lint-free cloth moistened with mild soapy water or diluted non-corrosive detergent. Dry the surfaces with a clean, soft cloth.
The cuff may be soaked in soapy water, rinsed and dried. However, never immerse the main device or allow liquid to enter its enclosure, sockets or electrical connections.
For disinfection, our manual recommends wiping the relevant surfaces twice for three minutes with clean gauze moistened with 70%–80% ethanol. Allow the surface to air-dry or remove the remaining disinfectant with a clean cloth. Keep ethanol away from flames and other ignition sources.
Patient-contact components should be disinfected after use by anyone with a known or suspected infection. Always follow the medical establishment’s infection-control policies.
We recommend regular inspections and annual calibration by a qualified organization. Please use replacement cuffs and other components supplied or authorized by us. Unapproved parts may affect measurement accuracy, device performance or safety.
FDA Clearance and Clinical Validation
The correct U.S. regulatory description for our DBP-01P is “FDA 510(k) cleared,” rather than “FDA approved.”
The official FDA Premarket Notification Database lists the DBP-01P and DBP-01HP under clearance number K231984. The FDA decision date was March 26, 2024.
The device is classified as a Class II noninvasive blood pressure measurement system under 21 CFR 870.1130, product code DXN. The FDA determined that the device was substantially equivalent to the identified legally marketed predicate device for its labeled intended use.
The FDA summary also includes clinical validation data obtained under ISO 81060-2:2018. The DBP-01P was evaluated in 85 subjects using its 17–42 cm wide-range cuff. The reported mean device–observer differences were 1.32 ± 3.18 mmHg for systolic pressure and 0.78 ± 3.15 mmHg for diastolic pressure. The results satisfied the protocol criteria described in the FDA submission.
Our product documentation also lists CE marking, ISO 13485 quality-management certification and ESH clinical validation. Customers should confirm the current certificates and destination-market configuration before procurement or deployment.
With proper installation, operation and maintenance, our DBP-01P provides a clear measurement workflow, wide cuff compatibility, guided arm positioning, voice assistance, automatic printing and flexible system-integration options. We designed it to support efficient, repeatable blood pressure screening in high-throughput professional healthcare environments.